Answering the Question: Why was Susvimo recalled?

4 min read

In October 2022, Genentech voluntarily recalled its Susvimo ocular implant, a novel treatment for wet age-related macular degeneration (AMD), just a year after its FDA approval. The recall was initiated after some implants failed quality performance standards during testing, providing the definitive answer to why was Susvimo recalled.

Quick Summary

The Susvimo ocular implant was recalled by Genentech in October 2022 due to a manufacturing defect in the septum, the seal that failed during repeated needle punctures. The issue was resolved through device updates and manufacturing improvements, allowing the product to be relaunched in 2024.

Key Points

  • The Cause: The Susvimo recall was triggered by a manufacturing flaw in the ocular implant's septum, which is the seal used for refilling.

  • Problem Detection: The defect was discovered during commercial supply testing, which showed some implants failed performance standards after being subjected to repeated needle punctures.

  • Recall Action: Genentech initiated a voluntary recall in October 2022, halting all new implantations and distribution of the device and insertion tool.

  • Existing Patients: Patients with existing implants were advised to continue refills as planned, as the recall did not necessitate removal of already implanted devices.

  • Resolution and Reintroduction: After resolving the manufacturing issue through device updates and production improvements, Genentech relaunched the updated Susvimo in July 2024.

  • FDA Approval: The reintroduction was granted with the approval of the U.S. Food and Drug Administration following a review of the updated device and manufacturing process.

  • Purpose: The recall, and its resolution, ultimately addressed issues with a device designed to reduce the treatment burden for patients with wet AMD by providing sustained medication release.

In This Article

The groundbreaking promise of Susvimo

Susvimo (ranibizumab injection) represented a significant advancement in the treatment of wet age-related macular degeneration (wet AMD), a leading cause of vision loss. Unlike standard-of-care eye injections that are often required monthly, Susvimo was a surgically implanted, refillable port delivery system designed to provide a continuous, sustained release of medication over several months. This innovative approach promised to reduce the frequent treatment burden for patients, who would need as few as two refills per year. The device, developed by Genentech (a member of the Roche Group), was first approved by the U.S. Food and Drug Administration (FDA) in 2021. Its introduction offered a new and promising alternative for patients seeking a less frequent dosing regimen.

The October 2022 voluntary recall

On October 18, 2022, Genentech initiated a voluntary recall of the Susvimo ocular implant, the insertion tool, and the initial fill kit in the U.S.. The recall was based on internal commercial supply testing that revealed a critical manufacturing defect in the device. As a result, Genentech halted all new implantations, including those in ongoing clinical trials. The company assured patients and physicians that there was no medical need to remove implants already in place and that refills could continue safely for those already implanted. This action demonstrated Genentech's commitment to patient safety, prioritizing product quality over market presence.

Identifying the root cause: A critical manufacturing flaw

The investigation into the implant's performance issues centered on the septum, which is the seal on the port delivery system used for refilling the medication. Genentech's CEO, Bill Anderson, explained that testing, which involved repeatedly puncturing the implants with a needle, showed that the seal could fail after repeated dosing. This manufacturing flaw led to septum dislodgement in some implants, potentially compromising the device's function and preventing proper refilling. While the reported rate of septum dislodgement was relatively low in clinical trials, the company’s internal reliability testing indicated that some implants did not meet their high-performance standards. The company acted decisively to address the issue before widespread problems could occur in the commercial setting.

Comparison of Susvimo implant versions and related recall actions

Feature Original Susvimo Implant (pre-2022) Updated Susvimo Implant (2024 release)
Recall Status Voluntarily recalled due to manufacturing defect. Reintroduced to the market following corrective actions.
Septum Quality Seal on the port was found to potentially fail after repeated needle punctures. Updated design with improved reliability standards based on further testing.
Refill Needle Original refill needle used with the recalled implant. Updated refill needle designed for improved performance with the new implant.
Manufacturing Process Initial process resulted in insufficient bonding of the septum. Corrective changes were implemented to improve manufacturing reliability.
Availability New implantations were paused from October 2022 until July 2024. Available again for new implantations and refills.

A path forward: Genentech's resolution and reintroduction

After the recall, Genentech dedicated time to addressing the manufacturing flaw. This involved making component-level updates to both the ocular implant and the refill needle. The company also implemented improvements to its manufacturing process to ensure the product met stringent performance standards. Following these corrective actions, and confirming the device's reliability through extensive testing, Genentech received approval from the FDA to reintroduce the updated Susvimo implant to the market. The reintroduction was formally announced in July 2024, signaling a return for this novel treatment option.

Navigating patient concerns and future considerations

During the recall period, physicians were advised to continue monitoring patients with existing implants and proceed with refills as needed. The reintroduction provides a renewed option for retina specialists and patients who are good candidates for this sustained-delivery system. The experience with Susvimo highlights the inherent challenges and complexities associated with developing and manufacturing innovative medical devices. The company's transparency and proactive measures to rectify the issue were noted by the medical community, reinforcing the importance of rigorous quality control in pharmacovigilance. The re-entry of Susvimo also paves the way for potential future advancements in sustained-release retinal treatments.

Conclusion: Lessons learned from the Susvimo recall

In summary, the reason why Susvimo was recalled was a specific manufacturing defect concerning the septum of the ocular implant, which compromised its reliability over multiple uses. Genentech's subsequent actions—halting new implantations, identifying the root cause, and developing and testing a new, more reliable device—demonstrated a commitment to patient safety and product quality. The successful reintroduction of Susvimo nearly two years later serves as a prime example of how pharmaceutical companies can effectively manage recalls, innovate through challenges, and restore confidence in their products. For patients with wet AMD, the return of Susvimo means access to a valuable treatment alternative, fulfilling its initial promise of reducing the treatment burden associated with frequent eye injections. For a detailed medical overview of the device, you can consult the official Genentech Susvimo documentation at https://www.susvimo.com/.

Frequently Asked Questions

Susvimo was recalled due to a manufacturing defect involving the implant's septum, a seal on the port delivery system. Internal testing revealed that the seal could fail after repeated needle punctures, leading to a voluntary recall.

The Susvimo implant is a refillable ocular implant used to deliver a continuous, sustained dose of ranibizumab for the treatment of wet age-related macular degeneration (wet AMD).

According to Genentech, there was no medical need for explantation (removal) for existing patients. Those with already implanted devices were advised to continue receiving refills as normal.

The voluntary recall was announced in October 2022. The updated Susvimo device was reintroduced to the market in July 2024, following FDA approval.

Following the recall, Genentech implemented component-level updates to the implant and refill needle. Improvements were also made to the manufacturing process to ensure the device met high performance standards.

The flaw was identified during internal quality control testing of the commercial supply. The tests, which simulated long-term use, showed that some implants did not perform to company standards when the septum was punctured repeatedly.

Yes, Genentech worked to resolve the manufacturing issues and made device updates. After being reviewed and approved by the FDA, the updated implant was relaunched in July 2024 for new implantations.